Keysight E1412A 6 1/2 Digit Multimeter Service Guide
21
Verification Tests
Verification Tests
Printing Results
You may want to modify the source code and recompile the project to add
statements to print the measured test results to your local printer. Manually you
can use the Performance Test Record as described on the previous page in the
section "Recording Your Test Results".
Test Frequency
See Table 4-1,
Recommended Test Equipment
, for test equipment requirements.
You should complete the Performance Verification tests at one year intervals. For
heavy use or severe operating environments, perform the tests more often.