Keysight E1412A 6 1/2 Digit Multimeter Service Guide
67
CALibration?
Adjustments
CALibration?
CALibration?
performs a calibration using the specified calibration value set by the
CALibration:VALue command and queries the calibration response to verify a
successful calibration.
Comments
–
Execution of this command begins the electronic adjustment for the
function and range the multimeter is set to. The adjustment is performed
based on the value stated in the CAL:VAL command and the multimeter
expects that value at the input terminals.
–
The command returns "0" to indicate there are no calibration errors and
the calibration was performed. A "1" is returned if a calibration error occurs
and a calibration is unable to be performed. The error message is reported
to the output buffer.
–
You must set CALibration:SECure:STATe OFF <code> to allow a calibration
to be performed. This requires that you know the calibration secure code.
The secure state enabled prevents unauthorized calibration of the
multimeter.
Calibrate the active function and range using the CAL:VALue:
CAL?
Perform the calibration.
monitor the status byte to detect calibration operation complete
enter statement
Enter cal response into computer to verify the calibration was successful.
Содержание E1412A
Страница 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Страница 2: ......
Страница 6: ...vi ...
Страница 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Страница 22: ...16 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Operating Instructions Warnings and Cautions ...
Страница 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Страница 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Страница 117: ...110 E1412A 6 1 2 Digit Multimeter Service Guide Keysight E1412A Multimeter Error Messages Execution Errors ...
Страница 118: ......