48
Keysight E1412A 6 1/2 Digit Multimeter Service Guide
Adjustments
Adjustment Procedures
–
A zero indicates that the adjustment was accepted.
–
A non-zero result indicates that an error has occurred. Send the
SYST:ERR? command and read the result. See Chapter 5 of the
E1412A
User’s Manual and SCPI Programmer’s Guide
for details.
Do not remove power from the mainframe during the multimeter’s
Zero Adjustment. You may lose ALL calibration memory if power is
removed while the multimeter is adjusting.
The automatic zero adjustment procedure takes about 5 minutes
to complete.
Figure 5-2
Zero Adjustment Test Setup.