Keysight E1412A 6 1/2 Digit Multimeter Service Guide
25
Performance Verification
Verification Tests
4
4. Select each function and range in the order shown in Table 4-1. Compare
the measurement results to the appropriate test limits shown in the table.
[1] For 2-wire ohms these errors assume you are using CALC:NULL:OFFS to null the effect of test lead resistance.
See the Keysight E1412A User’s Manual for the section titled 2-Wire Ohms Measurement Connections in
Chapter 1 and the section titled "NULL (Relative) Function" in Chapter 2. Add an additional 200 m
of error to
the specification value in the table if you do not null the lead resistance (e.g., 24 hour error for 100 mV range
would be ± 203 μV).
Q = Quick performance verification test points.
Table 4-2
Zero Offset Verification Test Points
INPUT
FUNCTION
QUICK
VERIFICATION
Keysight
E1412A
Range
Error from nominal (volts and current) Error
from actual (4-W and 2-W ohms)
24 hour [1]
90 day [1]
1 year [1]
4-wire short
DC Volt
Q
100 mV
± 3 μV
± 3.5 μV
± 3.5 μV
1 V
± 6 μV
± 7 μV
± 7 μV
10 V
± 40 μV
± 50 μV
± 50 μV
100 V
± 600 μV
± 600 μV
± 600 μV
300 V
± 5.4 mV
± 9 mV
± 9 mV
4-wire short
4-wire Ohms
and
2-wire Ohms
100
Ω
± 3 m
Ω
± 4 m
Ω
± 4 m
Ω
1 k
Ω
± 5 m
Ω
± 10 m
Ω
± 10 m
Ω
Q
10 k
Ω
± 50 m
Ω
± 100 m
Ω
± 100 m
Ω
100 k
Ω
± 500 m
Ω
± 1
Ω
± 1
Ω
1 M
Ω
± 10
Ω
± 10
Ω
± 10
Ω
10 M
Ω
± 100
Ω
± 100
Ω
± 100
Ω
100 M
Ω
± 10 k
Ω
± 10 k
Ω
± 10 k
Ω
open
DC Current
10 mA
± 1 μA
± 2 μA
± 2 μA
Q
100 mA
± 4 μA
± 5 μA
± 5 μA
1 A
± 60 μA
± 100 μA
± 100 μA
3 A
± 600 μA
± 600 μA
± 600 μA
Содержание E1412A
Страница 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Страница 2: ......
Страница 6: ...vi ...
Страница 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Страница 22: ...16 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Operating Instructions Warnings and Cautions ...
Страница 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Страница 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Страница 117: ...110 E1412A 6 1 2 Digit Multimeter Service Guide Keysight E1412A Multimeter Error Messages Execution Errors ...
Страница 118: ......