Keysight E1412A 6 1/2 Digit Multimeter Service Guide
Adjustments
Adjustment Procedures
7
Repeat steps 3 through 6 for the ranges and inputs given in the following
table.
DC Current Function Adjustment Points
Keysight E1412A Range
DC Standard Output
1
1 Valid calibration input values are 0.9 to 1.1 of Full Scale of the range to
be adjusted (9mA to 2A for the 3A range). For best results, set the
calibration value to full-scale (except the 3A range).
10 mA
10 mA
100 mA
100 mA
1 A
1 A
3 A
3 A
Figure 5-6
DC Current Gain Adjustment Setup.