:MEASure[:SCALar]:POWer[:AC] [<expected_value>[,<resolution>]]
[,(@3)]
This command measures power.
FUNCTION
DESCRIPTION
:MEASure[:SCALar]:VOLTage[:DC] <expected_value>
This command measures voltage.
FUNCTION
DESCRIPTION
default:
(@2)
response format:
<NR1>
<expected_value>
range:
-40 to +10 dBm
default:
0.00
resolution:
0.01
<resolution>
description:
<resolution> is supported only for compatibility with other instruments.
values:
0.01 dB | MIN | MAX | DEF
default:
0.01 dB
<source_list>
description:
Specifies which front-panel input is used for the measurement.
values:
DEF | (@3 )
default:
(@3)
response format:
<NR2> Power values are returned in dBm.
<expected_value>
range:
-50 to +50 vdc
default:
0.00
<source_list>