:ABORt Command
COMMAND
:ABORt
This command causes the instrument to abort, as quickly as possible, any
measurement in progress.
The
:ABORt
command is not complete until the current measurement is stopped.
The execution of an
ABORt
command sets false any Pending Operation Flags that
were set true by initiation of measuring.
COMMENTS
– If
:ABORt
is issued while the measurement cycle is idle (
:INIT:CONT OFF
and
pending operation flag is false), the command is ignored.
– If
:ABOR
t is issued while a single measurement is in progress, the
measurement is aborted and the pending operation flag is set false.
– If
:ABORt
is issued while repetitive measurements are being made (
:INIT:CONT
ON
), the measurement in progress is aborted, and the pending operation flag
is set false. Then, a new measurement is automatically initiated, and the
pending operation flag is set true.
– If
:ABORt
is issued while a block of measurements, such as an average, is in
progress, the measurement block is aborted, and the pending operation flag is
set false.
– When a measurement or block of measurements is aborted, the Measuring bit
in the Operation Status Register is set false.
– Aborting a measurement in progress invalidates the result.
RELATED
FRONT-PANEL
KEYS
Reset/Local