Programming
3
Keysight 53147A/148A/149A Operating and Programming Guide
229
Using :CONFigure with :READ?
The
:CONFigure
command causes the instrument to choose default settings for
the specified measurement.
:READ?
starts the measurement and queries the
result.
This sequence operates in the same way as the
:MEASure
query, but it allows you
to insert commands between
:CONFigure
and
:READ?
to specify a particular
setting.
For example, use
:CONF:FREQ 5 GHZ, 1HZ
to configure a default frequency measurement, where 1 Hz is the required
resolution and 5 GHz is the expected value.
Use
:READ?
to start the measurement and query the result.