Programming
3
Keysight 53147A/148A/149A Operating and Programming Guide
153
, provides instructions for
setting the GPIB address from the instrument’s front panel.
Changing the GPIB Address
1
Press and release the Shift key, and then press Menu(Reset/Local).
The instrument’s menu is displayed. Press the up or down arrow key repeatedly
until
GPIB ADDR
is displayed, and the LED indicator between the arrow keys
flashes. The current GPIB address is shown to the right of the blinking indicator
(>) flashes.
2
Press the right-arrow key. The blinking indicator changes direction (from
>
to
<
), and the current GPIB address blinks.
3
Press (or press and hold) the up-arrow or down-arrow key to change the GPIB
address (the available addresses are 0 to 30).
4
When your desired address is displayed, press the Enter key. The address you
selected is assigned, and the display returns to its normal operating mode. You
must
press the Enter key to complete the entry.
NOTE
Once the instrument is in Remote mode, all front-panel keys except the
Reset/
Local
key are disabled. As long as local-lockout is off, pressing the
Reset/Local
key returns the instrument to Local mode.
NOTE
To configure the instrument so that the a specific GPIB address is automatically
assigned each time you turn the instrument on, select the address, and then
save your current settings in SAV 0. The settings in SAV 0 are recalled each time
the instrument is turned on.