FRONT PANEL FAIL
X
The front panel or its interconnecting cable are
defective or not properly connected.
FPANEL FAIL
X
FPGA FAIL
X
X
A failure was detected in the FPGA
(Field Programmable Gate Array).
GPIB FAIL
X
X
A failure was detected in the GPIB hardware.
HETERODYNE PATH FAIL
X
A failure was detected in the heterodyne-path
circuit.
HET PATH FAIL
X
IIC FAIL
X
X
An attempt to write to the LCD display failed.
INSTCFG FAIL
X
X
The instrument’s configuration data is missing or
has become corrupted.
OVER TEMPERATURE
X
The instrument’s internal temperature is above the
acceptable limit.
OVERTEMP
X
PWR CAL FAIL
X
X
The instrument’s power-calibration data is missing
or has become corrupted.
POWER METER FAIL
X
A failure was detected in the Channel 2
power-measurement circuit.
PWR CKT FAIL
X
RAM FAIL
X
X
A failure was detected in RAM.
ROM FAIL
X
X
A failure was detected in ROM.
SAV SET FAIL
X
X
One or more of the sets of user settings is missing
or has become corrupted.
SERVICE FAIL
X
X
The instrument’s service data is missing or has
become corrupted.
THRU PATH FAIL
X
A failure was detected in the
through-path circuit of Channel 2.
THRU FAIL
X
VCO/COUNT CHAIN FAIL
X
A failure was detected in the VCO or the Count
Chain.
VCO/CNT FAIL
X
Table B-3
Error Messages (continued)