Programming
3
Keysight 53147A/148A/149A Operating and Programming Guide
251
:TRIGger Subsystem
COMMAND
:TRIGger[SEQuence | STARt]:HOLDoff . . . <numeric_value>
When
INIT:CONT
is
ON
, this command determines the rate at which
measurements are made by setting a delay between measurements. Its setting
corresponds to the front-panel rate setting as follows:
<NUMERIC_
VALUE> RATE
0.0 =
FAST
; 0.5 =
MEDium
; 1.0 =
SLOW
COMMENTS
– *RST: 0.0
– The only settings that are accepted are the three listed above.
– The measurement rate is also affected by the Resolution setting.
– When
INIT:CONT
is
OFF
, the holdoff has no effect.
– There is no setting for this command that corresponds to the front-panel rate
setting of
HOLD
. The
HOLD
mode is entered through
*RST
or
INIT:CONT OFF
.
The instrument then stops making measurements until it is triggered or until it
receives
INIT:CONT ON
.
RELATED
FRONT-PANEL
KEYS
Shift + Rate