77
Functions and Measurements
Making a Power Measurement with USB Power Sensor
7
Press
Meas Setup > Disp Range
to set the display range of the
measurement. when it is set to
On
, the display range will be
adjusted automatically according to the signal power
dynamic range. When it is set to Off, press
Top
and
Bottom
to
set the display range manually.
8
Press
General Setup > Freq
to set the power frequency for the
DUT. The current frequency will display in the bottom line.
9
Press
Power Range > Auto Power
to toggles the measurement
range between
On
and
Off
. When it is set to
On
, the power
range will be adjusted automatically. Use auto range when
you are not sure of the power level you are about to
measure. When it is set to
Off
, press
Range
to select the
upper range or lower range. The lower range uses the more
sensitive path and the upper range uses the attenuated path.
Table 1
Power Sensor Range
Sensor
Lower range
Upper range
U2000/1/2/4A
–60 dBm to –7 dBm
–7 dBm to +20 dBm
U2000/1/2H
–50 dBm to +3 dBm
+3 dBm to +30 dBm
U2000/1B
–30 dBm to +23 dBm
+23 dBm to +44 dBm
Содержание N9322C
Страница 8: ...Contents 6 ...
Страница 108: ...Functions and Measurements Files Operation 100 ...
Страница 174: ...Key Reference Trace 166 ...
Страница 191: ...183 Menu Maps BW Menu BW Menu BW RBW VBW VBW RBW Avg Type Log Pwr Power Voltage Auto Man Auto Man Auto Man ...
Страница 206: ...Menu Maps Name Editor Menu 198 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Страница 208: ...Menu Maps SPAN Menu 200 SPAN Menu Span Span Full Span Zero Span Last Span ...
Страница 210: ...Menu Maps Trace Det Menu 202 ...
Страница 213: ......
Страница 214: ...This information is subject to change without notice Keysight Technologies 2012 2014 Edition 1 July 2014 N9322 90002 ...