Functions and Measurements
Making a Power Measurement with USB Power Sensor
76
•
Before measuring low level signals—for example, 10 dB
above the lowest specified power for your power sensor.
Calibrating the power meter every time you cycle the power
on and off.
In most situations, you can press
Zero
to complete the two
steps (zero and cal) together.
Follow the steps below to make a basic average power
measurement.
1
Press
Preset
. (With
Preset Type
of
Factory
)
2
Press
Mode > Power Meter
to access the power meter mode.
3
Zero and calibrate the meter. Press
Zero
to make a Zero
operation of the power sensor followed by a calibration
operation.
4
Connect the power sensor to the USB
port
of N9322C. The
analyzer will recognize the power sensor automatically.
5
Connect the external attenuation, if required, with the power
sensor used.
6
Connect the analyzer to the base station. Connect the power
sensor to the signal to be measured. A typical BTS
connection is after the output of the power amplifier and
duplexer. The connection requires a cable and two Type- N
barrel connectors, in addition to the power sensor and
external attenuator.
CAUTION
To prevent damage to your sensor, do not exceed the power levels below.
U2000/1/2A +25 dBm (320 mW) average, 20 VDC
+33 dBm (2 W) peak, <10
μ
s
For more information on the maximum power of U2000 series power
sensors, please refer to the
Operating and Service Guide
of Keysight
U2000 Series USB Power Sensors.
Содержание N9322C
Страница 8: ...Contents 6 ...
Страница 108: ...Functions and Measurements Files Operation 100 ...
Страница 174: ...Key Reference Trace 166 ...
Страница 191: ...183 Menu Maps BW Menu BW Menu BW RBW VBW VBW RBW Avg Type Log Pwr Power Voltage Auto Man Auto Man Auto Man ...
Страница 206: ...Menu Maps Name Editor Menu 198 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Страница 208: ...Menu Maps SPAN Menu 200 SPAN Menu Span Span Full Span Zero Span Last Span ...
Страница 210: ...Menu Maps Trace Det Menu 202 ...
Страница 213: ......
Страница 214: ...This information is subject to change without notice Keysight Technologies 2012 2014 Edition 1 July 2014 N9322 90002 ...