45
Functions and Measurements
Measuring a Low-Level Signal
3
Set the center frequency, span and reference level:
Press
Frequency > 300 > MHz
Press
Span > 2 > MHz
Press
Amplitude > 40 > –dBm
.
4
Move the desired peak to the center of the display:
Press
Peak Search
Press
Marker > Marker To > Mkr -> CF
5
Reduce the span to 500 kHz, if necessary re- center the peak:
Press
Span > 500 > kHz
6
Set the attenuation to 20 dB:
Press
Amplitude > Attenuation > 20 > dB
Figure 3-10 A signal closer to the noise level
Note that increasing the attenuation moves the noise floor
closer to the signal level.
7
To allow more of the signal power to pass through the
analyzer, decrease the attenuation to 0 dB.