Getting Started
Helpful Tips
24
Amplitude Calibration
The analyzer privides the internal amplitude calibration
function. Please refer to the procedures below to perform an
amplitude calibration:
1
Press
[System] > More > Service > Calibration > Amp Alignment
2
2 Connect the CAL OUT and RF IN connector with a
N- BNC cable. Then Press
[Enter]
to continue.
The calibration will start immediately. The calibration will be
completed in serveral minutes.
Using an External Reference
To use an external 10 MHz source as the reference
frequency, connect the external reference source to the
REF
IN
connector on the rear panel. An
EREF
indicator will
display in the
left side
of the display. The signal level must
be in the range of –5 to +10 dBm. When the reference signal
is ready, the instrument switches the reference from time
base to the external reference.
Enable an Options
Option license key information is required to enable product
options. Contact your nearest Keysight office for purchasing
a license. Refer to the procedures below to activate the
options you have purchased.
1
Press
[System] > More > Service > Add Option
2
Enter the option number to be enabled. Press
[Enter]
to
confirm your input.
3
Enter the license key in the onscreen window. Press
[Enter]
to confirm your input and terminate the license key input.
4
Press
Activate License
. The option
will
be
enabled immediately.
The analyzer provides the trial license with limited time
usage(15/270/365 days) of the wanted options.
Содержание N9322C
Страница 8: ...Contents 6 ...
Страница 108: ...Functions and Measurements Files Operation 100 ...
Страница 174: ...Key Reference Trace 166 ...
Страница 191: ...183 Menu Maps BW Menu BW Menu BW RBW VBW VBW RBW Avg Type Log Pwr Power Voltage Auto Man Auto Man Auto Man ...
Страница 206: ...Menu Maps Name Editor Menu 198 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Страница 208: ...Menu Maps SPAN Menu 200 SPAN Menu Span Span Full Span Zero Span Last Span ...
Страница 210: ...Menu Maps Trace Det Menu 202 ...
Страница 213: ......
Страница 214: ...This information is subject to change without notice Keysight Technologies 2012 2014 Edition 1 July 2014 N9322 90002 ...