49
Functions and Measurements
Improving Frequency Resolution and Accuracy
Improving Frequency Resolution and Accuracy
This section provides information on using the frequency
counter function to improve frequency resolution and
accuracy.
1
Press
Preset
(With
Preset Type
of
Factory
)
2
Connect a cable from the front panel
CAL OUT
to
RF IN
;
Press
System > Service > Calibration > CAL OUT ON
to toggle on
and enable the 50 MHz amplitude reference signal.
3
Press
Frequency > Auto Tune
.
The analyzer will detect the signal peak and locate it to the
center of
the
display (Refer
to
).
4
Turn the frequency counter on:
Press
Marker > Function > Counter
.
5
Move the marker, with the front- panel knob, half- way down
the skirt of the signal response.
Figure 3-14 Using Frequency Counter
NOTE
The frequency and amplitude of the marker appears in the active function
area (this is not the counted result). The counted result appears in the
upper-right corner of the display to the right-side of
C1
.
Counted result
Содержание N9322C
Страница 8: ...Contents 6 ...
Страница 108: ...Functions and Measurements Files Operation 100 ...
Страница 174: ...Key Reference Trace 166 ...
Страница 191: ...183 Menu Maps BW Menu BW Menu BW RBW VBW VBW RBW Avg Type Log Pwr Power Voltage Auto Man Auto Man Auto Man ...
Страница 206: ...Menu Maps Name Editor Menu 198 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Страница 208: ...Menu Maps SPAN Menu 200 SPAN Menu Span Span Full Span Zero Span Last Span ...
Страница 210: ...Menu Maps Trace Det Menu 202 ...
Страница 213: ......
Страница 214: ...This information is subject to change without notice Keysight Technologies 2012 2014 Edition 1 July 2014 N9322 90002 ...