157
Key Reference
Sweep
Trigger
Accesses the menu of keys that allow you to select the
trigger mode of the analyzer. This softkey is only available in
zero span.
Free Run
New sweep starts as soon as possible after the current
sweep ends.
Video
Activates the trigger condition that starts the next sweep if
the detected RF envelope voltage rises to a level set by the
video trigger level. When
Video
is pressed, a line appears on
the display. The analyzer triggers when the input signal
exceeds the trigger level at the left edge of the display.
Change the trigger level by using the arrow keys, the knob,
or the numeric keypad.
External
Activates the trigger condition that will start a sweep when
the external voltage (connected to
EXT TRG IN
on the rear
panel) passes through approximately 2 volts. The external
trigger signal must be a 0 V to +5 V TTL signal.
Key access:
Trace > More
NOTE
This function is not available when the Resolution Bandwidth is less
than 1 kHz. Selecting a Resolution Bandwidth less than 1 kHz while in
Video Trigger mode, changes the Trigger mode to Free Run.
Содержание N9322C
Страница 8: ...Contents 6 ...
Страница 108: ...Functions and Measurements Files Operation 100 ...
Страница 174: ...Key Reference Trace 166 ...
Страница 191: ...183 Menu Maps BW Menu BW Menu BW RBW VBW VBW RBW Avg Type Log Pwr Power Voltage Auto Man Auto Man Auto Man ...
Страница 206: ...Menu Maps Name Editor Menu 198 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Страница 208: ...Menu Maps SPAN Menu 200 SPAN Menu Span Span Full Span Zero Span Last Span ...
Страница 210: ...Menu Maps Trace Det Menu 202 ...
Страница 213: ......
Страница 214: ...This information is subject to change without notice Keysight Technologies 2012 2014 Edition 1 July 2014 N9322 90002 ...