175
Instrument Messages
Device-Specific Errors
652
Connect CAL OUT to RF IN
you must connect the
CAL OUT
to
RF IN
with the appropriate
cable.
751
Instrument state set to initial values
While trying to load a trace or state, the state information
was found to be in error. This may be because the state had
been stored on a later revision of analyzer firmware. A
default set of state variables was loaded instead. There is
nothing wrong with the analyzer.
752
Unable to load state from file
Attempt to load a state from a file failed.
753
Unable to save state to file
Attempt to save a state to a file failed. See the associated
error messages for the cause.
757
Unable to load user state
A user preset failed, so the factory preset values were used.
Save a valid state into user preset and try again.
758
Unable to save user state
Attempt to save a user preset state failed.
760
Unable to query state
Query of state over the remote interface was unsuccessful.
762
Unable to load file
A failure occurred while loading a file.
764
Unable to save file
Attempt to save a file failed; the file was not saved.
Содержание N9322C
Страница 8: ...Contents 6 ...
Страница 108: ...Functions and Measurements Files Operation 100 ...
Страница 174: ...Key Reference Trace 166 ...
Страница 191: ...183 Menu Maps BW Menu BW Menu BW RBW VBW VBW RBW Avg Type Log Pwr Power Voltage Auto Man Auto Man Auto Man ...
Страница 206: ...Menu Maps Name Editor Menu 198 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Страница 208: ...Menu Maps SPAN Menu 200 SPAN Menu Span Span Full Span Zero Span Last Span ...
Страница 210: ...Menu Maps Trace Det Menu 202 ...
Страница 213: ......
Страница 214: ...This information is subject to change without notice Keysight Technologies 2012 2014 Edition 1 July 2014 N9322 90002 ...