4-14 Keysight N5224A/25A Service Guide
Troubleshooting
Power Up Troubleshooting
4-
Figure 4-3
Fan Power Cable Connections
3.
If the correct voltage is present at each connection and the fan connectors
are in good mechanical condition, suspect a defective fan. Refer to
“Removing and Replacing the Fans” on page 7-55
.
4.
If the correct voltage is not present, suspect a defective A18 system
“Removing and Replacing the A18 System
Troubleshooting LCD Display Problems
This procedure is intended to isolate the faulty assembly when the display is
dark. If the display is lit, but the color mix is faulty, refer to
1.
If the display is dim, the A3 display assembly is defective. Refer to
“Removing and Replacing the A1–A3 and Other Front Panel
.
There are no front panel adjustments for intensity and contrast of the LCD.
Содержание N5224A
Страница 124: ...4 52 Keysight N5224A 25A Service Guide Troubleshooting Instrument Block Diagrams Sheet 2 4 ...
Страница 132: ...5 8 Keysight N5224A 25A Service Guide Theory of Operation Synthesized Source Group Operation 5 Figure 5 3 Source Group ...
Страница 149: ...Keysight N5224A 25A Service Guide 5 25 Theory of Operation Receiver Group Operation Figure 5 6 4 Port Receiver Group ...
Страница 150: ...5 26 Keysight N5224A 25A Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 7 2 Port Receiver Group ...
Страница 266: ...6 110 Keysight N5224A 25A Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 346: ...8 18 Keysight N5224A 25A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...