Keysight N5224A/25A Service Guide 3-43
Tests and Adjustments
Adjustments
Procedure
1.
Connect the equipment as shown in
. Connect a GPIB cable
between the network analyzer and the power meter.
Figure 3-15
Equipment Setup for the Source Adjustment
2.
.Press UTILITY
, then
, then
, then click
Source Adjustment.
3.
Ensure the GPIB settings are correct.
4.
Follow the instructions and prompts as they are displayed.
IF Gain Adjustment
The IF gain adjustment is used to adjust the IF gain of the network analyzer.
Procedure
1.
Press UTILITY
, then
, then
, then click IF
Gain Adjustment.
2.
Follow the instructions and prompts as they are displayed.
Receiver Characterization
This characterizes the receivers in your analyzer.
Procedure
1.
Press UTILITY
, then
, then
, then click
Receiver Characterization.
System
Service
Adjustments
System
Service
Adjustments
System
Service
Adjustments
Содержание N5224A
Страница 124: ...4 52 Keysight N5224A 25A Service Guide Troubleshooting Instrument Block Diagrams Sheet 2 4 ...
Страница 132: ...5 8 Keysight N5224A 25A Service Guide Theory of Operation Synthesized Source Group Operation 5 Figure 5 3 Source Group ...
Страница 149: ...Keysight N5224A 25A Service Guide 5 25 Theory of Operation Receiver Group Operation Figure 5 6 4 Port Receiver Group ...
Страница 150: ...5 26 Keysight N5224A 25A Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 7 2 Port Receiver Group ...
Страница 266: ...6 110 Keysight N5224A 25A Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 346: ...8 18 Keysight N5224A 25A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...