Keysight N5224A/25A Service Guide 3-23
Tests and Adjustments
System Verification
Figure 3-9
NIST Traceability Path for Calibration and Verification Standard
Performing System Verification
The following verification procedure is automated by the analyzer firmware.
The process for the verification is:
— connect cables to the analyzer test ports
— perform a calibration or recall a recent calibration
— run the system verification program for the verification devices
Each time through the verification process, you are prompted to make
necessary connections and perform or recall a calibration as part of performing
the verification. If you select to perform a calibration, you are guided through
the calibration procedure. This part of the process can be eliminated if you