3-46 Keysight N5224A/25A Service Guide
Tests and Adjustments
Adjustments
3-
Figure 3-18
Equipment Setup 3 for the Receiver Adjustment
EE Default Adjustment
This sets the EEPROM data to their default values.
Procedure
1.
Press UTILITY
, then
, then
, then click EE
Default Adjustment.
2.
On the dialog box, select Initialize rather than Adjust/Verify because an
adjustment is typically unnecessary.
3.
Follow the instructions and prompts as they are displayed.
IF Response Adjustment (Option 090, 093, or 094 Only)
The IF Response adjustment is required when upgrading to Option 090, 093
and or 094. In addition, this adjustment must be performed each time a new
configuration is set up or if cables are changed:
Procedure
1.
Press UTILITY
, then
, then
, then click IF
Response Adjustment.
2.
Follow the instructions and prompts as they are displayed.
System
Service
Adjustments
System
Service
Adjustments
Содержание N5224A
Страница 124: ...4 52 Keysight N5224A 25A Service Guide Troubleshooting Instrument Block Diagrams Sheet 2 4 ...
Страница 132: ...5 8 Keysight N5224A 25A Service Guide Theory of Operation Synthesized Source Group Operation 5 Figure 5 3 Source Group ...
Страница 149: ...Keysight N5224A 25A Service Guide 5 25 Theory of Operation Receiver Group Operation Figure 5 6 4 Port Receiver Group ...
Страница 150: ...5 26 Keysight N5224A 25A Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 7 2 Port Receiver Group ...
Страница 266: ...6 110 Keysight N5224A 25A Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 346: ...8 18 Keysight N5224A 25A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...