Replacing 6000L Assemblies
5
6000L Series Oscilloscopes Service Guide
75
Figure 15
Disconnecting ribbon cable and releasing tab retainers
6
Pull the front panel off of the oscilloscope.
7
Lift the front panel PC board to remove it from the front panel.
8
Reverse this procedure to replace the front panel.
Front Panel
Ribbon Cable
Retainer Tab
System Board
Front
Panel
Retainer Tab Slot
Retainer Tab
Center Retainer Tab
100 MHz Auto
Probe Label
500 MHz Auto
Probe Label
Front
Panel PC
Board