44
6000L Series Oscilloscopes Service Guide
2
Testing Performance
Figure 7
Connect equipment for external trigger sensitivity test
2
Set up the oscilloscope.
a
Press the
Save/Recall
key, then press the
Default Setup
softkey.
3
Change the signal generator output frequency as follows:
•
For 500 MHz and 1 GHz models, set the signal generator output frequency
to 500 MHz
•
For 100 MHz models, set the signal generator output frequency to 100 MHz
Oscilloscope
Signal
Generator
Power Meter
Power Splitter
Power Sensor
N to BNC adapter
N Cable
50
Ω
Feedthrough
*
*
Required for all models.