Testing Performance
2
6000L Series Oscilloscopes Service Guide
43
2
Verify the trigger sensitivity at maximum bandwidth.
1 GHz models: 1 GHz
500 MHz models: 500 MHz
100 MHz models: 100 MHz
a
Press the
Save/Recall
key, then press the
Default Setup
softkey.
b
On 500 MHz and 1 GHz models, set channel 1
Imped
to
50 Ohm
.
c
Set the output frequency of the signal generator to the maximum bandwidth
of the oscilloscope and set the amplitude to about 10 mV
pp
.
d
Press the
AutoScale
key.
e
Set the time base to 10 ns/div.
f
Set channel 1 to 5 mV/div.
g
Decrease the amplitude from the signal generator until 1 vertical division of
the signal is displayed.
The trigger is stable when the displayed waveform is stable. If the trigger is
not stable, try adjusting the trigger level. If adjusting the trigger level makes
the trigger stable, the test still passes. If adjusting the trigger does not help,
see the “Troubleshooting” chapter. Then return here.
h
Record the result as Pass or Fail in the Performance Test Record (see
).
3
Repeat this procedure for the remaining oscilloscope channels.
Test External Trigger Sensitivity
The following test is optional because External Trigger Sensitivity is a
characteristic, not a specification.
Verify the external trigger sensitivity at these settings:
•
1 GHz and 500 MHz bandwidth models:
500 MHz, 500 mV
pp
•
100 MHz bandwidth models:
100 MHz, 500 mV
pp
1
Connect the equipment (see
a
Use the N cable to connect the signal generator to the power splitter input.
b
Connect one output of the power splitter to the Aux Trig input through a 50
Ω
feedthrough termination.
c
Connect the power sensor to the other output of the power splitter.
Содержание InfiniVision 6000L Series
Страница 1: ...Keysight InfiniVision 6000L Series Oscilloscopes Service Guide...
Страница 8: ...8 6000L Series Oscilloscopes Service Guide Contents...
Страница 10: ...10 6000L Series Oscilloscopes Service Guide Figures...
Страница 12: ...12 6000L Series Oscilloscopes Service Guide Tables...
Страница 48: ...48 6000L Series Oscilloscopes Service Guide 2 Testing Performance...
Страница 54: ...54 6000L Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Страница 86: ...86 6000L Series Oscilloscopes Service Guide 5 Replacing 6000L Assemblies...
Страница 94: ...94 6000L Series Oscilloscopes Service Guide 6 Replaceable Parts...