Replacing 6000L Assemblies
4
Replace the fan.
a
Place the rectangular part of the flexible fan mount over the fan.
b
Place the pegs in the flexible fan mount into the holes in the top and bottom
of the fan.
c
Align the holes in the bottom of the flexible fan mount over the studs on the
chassis as you slide the fan over the divider in the chassis.
d
Ensure that the notch in the rectangular portion of the fan mount aligns with
the front of the chassis.
Figure 20