34
6000L Series Oscilloscopes Service Guide
2
Testing Performance
1
Connect the equipment (see
a
Use the N cable to connect the signal generator to the input of the power
splitter input.
b
Connect the power sensor to one output of the power splitter.
c
For 500 MHz and 1 GHz oscilloscope models use an N-to-BNC adapter to
connect the other splitter output to the channel 1 input.
d
For 100 MHz oscilloscopes, use an N-to-BNC adapter and 50
Ω
feedthrough
termination to connect the other splitter output to the channel 1 input on
the oscilloscope.
Figure 5
Connect equipment for bandwidth test
2
Set up the power meter.
Oscilloscope
Signal
Generator
Power Meter
Power Splitter
Power Sensor
N to BNC adapter
N Cable
50
Ω
Feedthrough
*
*
100 MHz models only.