Testing Performance
2
6000L Series Oscilloscopes Service Guide
41
To verify trigger sensitivity
This test verifies the trigger sensitivity. In this test, you will apply a sine wave to
the oscilloscope at the upper bandwidth limit. You will then decrease the
amplitude of the signal to the specified levels, and check to see if the oscilloscope
is still triggered.
The internal trigger sensitivity test is mandatory because it is a specification. The
external trigger test is optional because it is a characteristic, not a specification.
Test limits for:
• Internal trigger:
•
< 10 mV/div: greater of 1 div or 5 mV
pp
>=10 mV/div: 0.6 div
• External trigger:
1 GHz and 500 MHz models:
(DSO6104L/6054L)
DC to 500 MHz: < 500 mV
pp
100 MHz models:
(DSO6014L)
DC to 100 MHz: < 500 mV
pp
Table 11
Equipment Required to Verify Trigger Sensitivity
Equipment
Critical Specifications
Recommended Model/Part
Signal Generator
25-MHz, 100-MHz, 500-MHz, and
1 GHz sine waves
Keysight N5181A, or E4400B, or
8648A
Power splitter
Outputs differ < 0.15 dB
Keysight 11667A
Cable
BNC, Qty 3
Keysight 10503A
Adapter
N (m) to BNC (f), Qty 3
Keysight 1250-0780
Feedthrough
50
Ω
BNC (f) to BNC (m) Qty 2
Keysight 0960-0301
Содержание InfiniVision 6000L Series
Страница 1: ...Keysight InfiniVision 6000L Series Oscilloscopes Service Guide...
Страница 8: ...8 6000L Series Oscilloscopes Service Guide Contents...
Страница 10: ...10 6000L Series Oscilloscopes Service Guide Figures...
Страница 12: ...12 6000L Series Oscilloscopes Service Guide Tables...
Страница 48: ...48 6000L Series Oscilloscopes Service Guide 2 Testing Performance...
Страница 54: ...54 6000L Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Страница 86: ...86 6000L Series Oscilloscopes Service Guide 5 Replacing 6000L Assemblies...
Страница 94: ...94 6000L Series Oscilloscopes Service Guide 6 Replaceable Parts...