Testing Performance
2
6000L Series Oscilloscopes Service Guide
33
To verify bandwidth
This test checks the bandwidth of the oscilloscope. In this test you will use a
signal generator and a power meter.
1 GHz Models
Test limits at 2 mV/div to 5 V/div
• All channels (-3 dB)
•
dc to 1 GHz
500 MHz Models
Test limits at 2 mV/div to 5 V/div
• All channels (-3 dB)
•
dc to 500 MHz
100 MHz Models
Test limits at 1 mV/div to 5 V/div
• All channels (-3 dB)
•
dc to 100 MHz
Table 8
Equipment Required to Verify Bandwidth
Equipment
Critical Specifications
Recommended Model/Part
Signal Generator
100 kHz - 1 GHz at 200 mVrms
Keysight N5181A, or E4400B,
or 8648A
Power Meter/Sensor
1 MHz - 1 GHz ±3% accuracy
Keysight E4418B/8482A
Power Splitter
outputs differ by < 0.15 dB
Keysight 11667A
Cable
Type N (m) 24 inch
Keysight 11500B
Adapter
Type N (m) to BNC (m)
Keysight 1250-0082
Feedthrough
50
Ω
BNC (f) to BNC (m)
Keysight 0960-0301
Содержание InfiniVision 6000L Series
Страница 1: ...Keysight InfiniVision 6000L Series Oscilloscopes Service Guide...
Страница 8: ...8 6000L Series Oscilloscopes Service Guide Contents...
Страница 10: ...10 6000L Series Oscilloscopes Service Guide Figures...
Страница 12: ...12 6000L Series Oscilloscopes Service Guide Tables...
Страница 48: ...48 6000L Series Oscilloscopes Service Guide 2 Testing Performance...
Страница 54: ...54 6000L Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Страница 86: ...86 6000L Series Oscilloscopes Service Guide 5 Replacing 6000L Assemblies...
Страница 94: ...94 6000L Series Oscilloscopes Service Guide 6 Replaceable Parts...