Testing Performance
2
6000L Series Oscilloscopes Service Guide
31
Figure 3
Connect equipment for voltage measurement accuracy test
4
Adjust the output so that the multimeter reading displays the first Volts/div
calibrator setting value in
.
Wait a few seconds for the measurement to settle.
5
Press the
Y2
softkey, then position the Y2 cursor to the center of the voltage
trace using the Entry knob.
The
Δ
Y value on the lower line of the display should be within the test limits
. If a result is not within the test limits, go to the “Troubleshooting”
chapter. Then return here.
6
Continue to check the voltage measurement accuracy with the remaining
Volts/div setting values in
.
Oscilloscope
Oscilloscope
Calibrator
Digital
Multimeter
BNC Tee
BNC (f) to dual
banana
Содержание InfiniVision 6000L Series
Страница 1: ...Keysight InfiniVision 6000L Series Oscilloscopes Service Guide...
Страница 8: ...8 6000L Series Oscilloscopes Service Guide Contents...
Страница 10: ...10 6000L Series Oscilloscopes Service Guide Figures...
Страница 12: ...12 6000L Series Oscilloscopes Service Guide Tables...
Страница 48: ...48 6000L Series Oscilloscopes Service Guide 2 Testing Performance...
Страница 54: ...54 6000L Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Страница 86: ...86 6000L Series Oscilloscopes Service Guide 5 Replacing 6000L Assemblies...
Страница 94: ...94 6000L Series Oscilloscopes Service Guide 6 Replaceable Parts...