Maintenance should be performed periodically to keep the instrument in good
condition.
Cleaning
Before performing cleaning, turn off the instrument, and disconnect power cable
from the rear panel. Use a dry cloth to clean the external case parts.
To prevent electrical shock, do not perform cleaning when the instrument is turned
on, and do not use a wet cloth.
Self-test and Diagnosis
This instrument provides the following functions to check the operation. Perform
the following functions as necessary. Open the measurement terminals to perform
self-test, self-calibration, or diagnosis.
•
Self-test
•
Self-calibration
•
Diagnosis
If problems arise, contact Keysight Technologies.
NOTE
Interlock Open/Close test and High Voltage LED test
Interlock function limits the maximum output voltage to ± 42 V when the
measurement terminal is touchable. For safety, this function must be checked before
using the instrument after power on at least once a day by performing the diagnosis.
Calibration
Calibration and adjustments must be performed periodically so that the instruments
satisfy the specifications, and keep a good condition. It is recommended to perform
the calibration once a year at least. For the calibration and adjustments, contact
Keysight Technologies. Trained service personnel will perform the calibration and
adjustments.