1- 12
Keysight B1507A User’s Guide, Edition 1
Introduction
Test Fixture Front View
Test Fixture Front View
This section describes the front view of the test fixture and selector.
1. Fixture cover
The fixture cover should be closed to avoid electrical shock by touching
measurement terminals and to prevent a device under test from external noise.
When the fixture cover is open, maximum output voltage is limited to ± 42 V.
WARNING
Hazardous voltage, instrument maximum output voltage may appear at the
Force, Sense, and Guard terminals if the fixture cover is closed.
WARNING
Make sure that the cover is closed properly before starting measurement. Do
not perform the measurement when a wire is protruding from the fixture cover.
Assurez-vous que le couvercle est fermé correctement avant de commencer la
mesure. Ne pas effectuer la mesure lorsqu’un câble dépasse du couvercle de
l'appareil.
2. 3-pin inline package socket module
Attach the module to the measurement terminals if your device under test (DUT)
is a 3-pin inline packaged device.
3. Measurement terminals
Connects the connection wire or attaches the socket module for connecting your
DUT. For more information, see “Measurement Terminals” on page 1-13.
1
2
Test fixture
6
3
5
4
Selector
7