1- 10
Keysight B1507A User’s Guide, Edition 1
Introduction
Mainframe Rear View
4. GPIB interface
Use an Keysight 82357A USB/GPIB interface or Keysight 10833A/B/C/D
GPIB cable to connect to an external computer or equipment.
5. Video output terminal
VGA connector. For an external display. Signal to the built-in LCD is also
applied to this terminal.
6. Measurement Resources
See A to D described later.
7. Circuit Common (
) and frame ground (
) terminals
Normally, connect the terminals together by using the shorting bar. For floating
measurement, remove the shorting bar.
WARNING
If the Circuit Common terminal is
not
connected to the frame ground terminal
(for floating measurement), a potential shock hazard may present. Do not
touch any of measurement circuit at any time while a floating measurement is
in progress.
CAUTION
For floating measurement, do not apply dangerous voltage to the Circuit Common
terminal. Failure to heed this caution may result in damage to the instrument.
8. Zero Check terminal
Ground reference point of the instrument.
CAUTION
The Zero Check terminal can be used for the service purpose only. For the normal
operation, leave this terminal open and do not connect anything to this terminal.
Connecting anything can damage the instrument.
9. GNDU terminal
0 V constant voltage source. Used for the reference of measurement ground.
Triaxial connector.
10. LAN interface
RJ45 connector.
11. USB interfaces
USB, 2 ports. For keyboard, mouse, or peripherals.