Keysight B1507A User’s Guide, Edition 1
3- 7
How To Perform Measurement
How To Perform Capacitance Measurement
For more information on setting items, see “Setup” on page 4-11.
Step 6.
Click the Start Calibration... button in the toolbar to measure the calibration data of
measurement path.
Follow the instruction displayed on the dialog to measure the open and short
calibration data.
If you have measured and saved the calibration data of measurement path, click the
Select Calibration Data... button in the toolbar to select and load the stored
calibration data.
Step 7.
Set your DUT on the test fixture and click the Start Measurement button
to
start a measurement.
You can stop the measurement by clicking the Stop Measurement button
during a measurement execution.
NOTE
If your DUT is diode, connect the cathode of your DUT to the Collector/Drain
terminal of the test fixture and the anode of your DUT to the Emitter/Source
terminal of the test fixture.
Step 8.
Click the Characteristics tab to display the Characteristics area.
After a measurement is finished, a measured capacitance characteristics curve is
displayed on the Characteristics area. This example shows the Ciss, Coss, and Crss,
characteristics curves.