Keysight B1507A User’s Guide, Edition 1
1- 27
Introduction
Measurement Resources
Table 1-6
MPSMU Measurement Voltage Values and Resolutions
Table 1-7
MPSMU Measurement Current Values and Resolutions
Range
Measurement
Value
a
Measurement Resolutions
High Speed ADC
High Resolution ADC
0.5 V
0
|V|
0.55 V
25
V
0.5
V
2 V
0
|V|
2.2 V
100
V
2
V
5 V
0
|V|
5.5 V
250
V
5
V
20 V
0
|V|
22 V
1 mV
20
V
40 V
0
|V|
44 V
2 mV
40
V
100 V
0
|V|
100 V
5 mV
100
V
Range
Measurement
Value
a
a. This column is applied to the auto ranging or the limited auto ranging.
For fixed ranging, maximum measurement value is
Range
column value.
Measurement Resolutions
High Speed ADC
High Resolution ADC
1 nA
0
|I|
1.15 nA
50 fA
10 fA
10 nA
0
|I|
11.5 nA
500 fA
10 fA
100 nA
0
|I|
115 nA
5 pA
100 fA
1
A
0
|I|
1.15
A
50 pA
1 pA
10
A
0
|I|
11.5
A
500 pA
10 pA
100
A
0
|I|
115
A
5 nA
100 pA
1 mA
0
|I|
1.15 mA
50 nA
1 nA
10 mA
0
|I|
11.5 mA
500 nA
10 nA
100 mA
0
|I|
100 mA
5
A
100 nA
Содержание B1507A