3- 4
Keysight B1507A User’s Guide, Edition 1
How To Perform Measurement
How To Perform Capacitance Measurement
Preparing Measurement
WARNING
Set the instrument output off before connecting or disconnecting connection
wire.
Press the mainframe front panel Stop key to set the source output off. And
confirm that the mainframe front panel High Voltage indicator is not lit.
WARNING
To prevent electrical shock and DUT damage, do not connect or disconnect the
DUT while the instrument is applying voltage or current.
When you touch the DUT after measurement, devise a countermeasure of
residual charge and heat to prevent electrical shock and burn. Use glove and
any tool. Also have enough time for discharge and radiation.
Step 1.
Prepare the test fixture and other accessories to connect your device under test
(DUT).
For more information on the test fixture and its accessories, see the following
sections:
•
“To Connect Selector and Test Fixture” on page 2-11
•
“3-pin Inline Package Socket Module” on page 2-17
•
“Accessories for Connecting a DUT” on page 2-18
Step 2.
Connect the test fixture and, 3-pin inline package socket module or other accessories
that is adequate for your DUT.
For more information, see the following sections:
•
“To Connect Selector and Test Fixture” on page 2-11
•
“3-pin Inline Package Socket Module” on page 2-17
•
“Accessories for Connecting a DUT” on page 2-18
Step 3.
Turn the instrument on.
Don’t connect your DUT to the test fixture at this time.