Keysight B1505A Configuration and Connection Guide
2-17
N1259A Connection Guide
Output Connection
To Use Universal R-Box
The universal R-box is a blank box, kind of a do-it-yourself kit for installing your desired
resistor.
To install the resistor, see the following simple instruction.
•
Required parts:
your desired resistor, adequate rating
hexlobe (torx type) screwdriver T-10
label, for noting the resistance value
•
Instruction:
1. Remove the cover bottom of the blank box.
2. Solder the resistor between the terminals 1 and 2.
To prevent discharge and any accident, make enough space between the resistor
leads and the shield or chassis. For example, make space of about 6 mm from the
leads for maximum 3000 V output.
3. Reattach the cover.
Make enough space between the resistor leads and the cover.
4. Write the resistance value in the label.
5. Affix the label to the top cover.
Figure 2-8
To Install Resistor in the N1259A-035 Universal R-Box
Содержание B1505A
Страница 1: ...Keysight Technologies B1505A Power Device Analyzer Curve Tracer Configuration and Connection Guide ...
Страница 3: ......
Страница 9: ...1 Configuration Guide ...
Страница 83: ...2 N1259A Connection Guide ...
Страница 107: ...3 N1265A Connection Guide ...
Страница 132: ...3 26 Keysight B1505A Configuration and Connection Guide N1265A Connection Guide Output Connection ...
Страница 133: ...4 N1272A and N1273A Connection Guide ...
Страница 148: ...4 16 Keysight B1505A Configuration and Connection Guide N1272A and N1273A Connection Guide Output Connection ...
Страница 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Страница 203: ...6 Accessory Dimensions ...
Страница 219: ...7 Connection and Ordering Examples ...
Страница 296: ...7 78 Keysight B1505A Configuration and Connection Guide Connection and Ordering Examples Upgrading from existing B1505A ...
Страница 297: ......