Keysight B1505A Configuration and Connection Guide
5-45
Connection Guide for Wafer Prober and Your Own Test Fixture
To Connect Gate Charge Adapter
To Connect N1275A
The N1275A is used to enable on-wafer gate charge measurements with the UHCU (max.
500 A) and the B1513B/C HVSMU (max. 3 kV). You need to switch signal paths of IV
measurement and gate charge measurement manually. And also, you need manual
switching between high voltage and high current gate charge measurements. The adapter
output can be extended to your probe station by using the N1254A-524 prober system
cable. (Note that the B1513A cannot be used for the gate charge measurement.)
You can also mount current control device or load resistor for the gate charge measurement
on the top of this adapter.
The N1275A should be installed in the shielding box with your DUT interface. Connection
example is shown in
. For the color-codes of cables and connectors in this
figure, see
.
Table 5-36
To Connect N1275A Input
Table 5-37
To Connect N1275A Output
N1254A terminals
Required cables Input
terminals
High
Force
N1275-61701 Ultra high current banana cable, 30 cm
(furnished in the N1275A)
High
Force
From
N1254A
Low
Sense
B1512-61741 BNC cable, 30 cm (furnished in the
N1275A)
Low
Sense
From
N1254A
B1505A terminals
Required cables
Input terminals
MCSMU
Force
16494A Triaxial cable, 1.5 m, 1 ea.
Current
Control
MCSMU
Force
Sense
16494A Triaxial cable, 1.5 m, 1 ea.
Sense
Output terminals
Required cables/Description
Connect to
High Force
To
DUT
Ultra high current cable with banana(plug). N1254A-523
can be used.
DUT high terminal
(ex: drain)
Low Sense
To
DUT
Coaxial cable with BNC(m) connector. N1254A-503 can
be used.
DUT low terminal
(ex: source)
To DUT
To DUT
From
N1254A Opt 524
From
N1254A Opt 524
Signal
Circuit
Common
Low Sense
±10 V
Max
High Force
±60 V
Max
Sense
Force
From B1505A
±30 V
Max
Current Control
MCSMU
Содержание B1505A
Страница 1: ...Keysight Technologies B1505A Power Device Analyzer Curve Tracer Configuration and Connection Guide ...
Страница 3: ......
Страница 9: ...1 Configuration Guide ...
Страница 83: ...2 N1259A Connection Guide ...
Страница 107: ...3 N1265A Connection Guide ...
Страница 132: ...3 26 Keysight B1505A Configuration and Connection Guide N1265A Connection Guide Output Connection ...
Страница 133: ...4 N1272A and N1273A Connection Guide ...
Страница 148: ...4 16 Keysight B1505A Configuration and Connection Guide N1272A and N1273A Connection Guide Output Connection ...
Страница 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Страница 203: ...6 Accessory Dimensions ...
Страница 219: ...7 Connection and Ordering Examples ...
Страница 296: ...7 78 Keysight B1505A Configuration and Connection Guide Connection and Ordering Examples Upgrading from existing B1505A ...
Страница 297: ......