4-10
Keysight B1505A Configuration and Connection Guide
N1272A and N1273A Connection Guide
Output Connection
5. Close the fixture cover and perform measurement.
WARNING
Hazardous voltage, instrument maximum output voltage may appear at the
measurement terminals if the fixture cover is closed.
WARNING
Open the Guard terminal of the test fixture. Do not connect anything to it.
NOTE
A short plate is furnished with the socket module. It is used for performing the short
correction before the capacitance measurement. Set it on the socket before the short
correction and remove it after the correction is completed.
Socket module
Test fixture
DUT
Chassis
common
Содержание B1505A