Keysight B1505A Configuration and Connection Guide
4-13
N1272A and N1273A Connection Guide
Output Connection
The following shows the positions of the upside terminals in the blank module.
7. Connect the other end of the cables to the terminals on the back side of the socket
board.
If a socket terminal does not have a screw or nut for securing a lug terminal of the
cable, prepare it. Ensure the terminal position of the blank module and the socket
board, and connect the cables correctly.
8. Fix the bottom cover to the blank module using a T10 Torx screwdriver.
CAUTION
Do not apply voltage/current over the maximum limit of the socket module.
Terminal rated voltage: 3000 Vd c
Terminal rated current: 120 mA
Collector/Drain
Base/Gate
Emitter/Source
AC/DC Guard
Blank Module
(Wiring side view)
Содержание B1505A
Страница 1: ...Keysight Technologies B1505A Power Device Analyzer Curve Tracer Configuration and Connection Guide ...
Страница 3: ......
Страница 9: ...1 Configuration Guide ...
Страница 83: ...2 N1259A Connection Guide ...
Страница 107: ...3 N1265A Connection Guide ...
Страница 132: ...3 26 Keysight B1505A Configuration and Connection Guide N1265A Connection Guide Output Connection ...
Страница 133: ...4 N1272A and N1273A Connection Guide ...
Страница 148: ...4 16 Keysight B1505A Configuration and Connection Guide N1272A and N1273A Connection Guide Output Connection ...
Страница 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Страница 203: ...6 Accessory Dimensions ...
Страница 219: ...7 Connection and Ordering Examples ...
Страница 296: ...7 78 Keysight B1505A Configuration and Connection Guide Connection and Ordering Examples Upgrading from existing B1505A ...
Страница 297: ......