2-22
Keysight B1505A Configuration and Connection Guide
N1259A Connection Guide
Output Connection
N1259A-013 Curve Tracer Test Adapter Socket Module
This module provides a socket available for connecting a test adapter designed for
connecting to Tektronix 370 series curve tracers. Socket module internal connection is
shown in
To use this module, see the following simple instruction.
•
Required parts:
N1254A-508 or N1254A-509 connection wire, 1 ea. for one connection
•
Instruction:
1. Attach the socket module to the test fixture.
2. Connect your test adapter to the socket.
3. Connect wires between the socket module terminals and the fixture output
terminals.
For making the Kelvin connection, Force and Sense must be connected to Force
and Sense of the socket module respectively.
If MCSMU, HCSMU, or DHCSMU is used, High Force and Sense must be
connected to Force and Sense used for the high terminal of a device under test
(DUT), respectively. Low Force and GNDU Force must be connected to Force used
for the low terminal of the DUT, and Low Sense and GNDU Sense must be
connected to Sense used for the low terminal of the DUT. See
example.
4. Set the DUT on your test adapter.
5. Close the fixture cover and perform measurement.
CAUTION
Do not apply voltage/current over the maximum limit of the socket module.
Figure 2-13
Curve Tracer Test Adapter Socket Module
Collector / Drain
Emier / Source
Base / Gate
N1259A Opt 013
Force Sense
Force Sense
Force Sense
Curve Tracer Test Adapter Socket
1
2
5
6
4
3
Internal connection
1: Collector/Drain Force
2: Collector/Drain Sense
3: Emitter/Source Force
4: Emitter/Source Sense
5: Base/Gate Force
6: Base/Gate Sense
A
B
C
Maximum voltage: 3000 V
Maximum current:
A-Force: 500 A pulse, 39 A dc
A-Sense: 40 A pulse, 2 A dc
B-Force: 500 A pulse, 39 A dc
B-Sense: 40 A pulse, 2 A dc
C-Force: 40 A pulse, 2 A dc
C-Sense: 40 A pulse, 2 A dc
Содержание B1505A
Страница 1: ...Keysight Technologies B1505A Power Device Analyzer Curve Tracer Configuration and Connection Guide ...
Страница 3: ......
Страница 9: ...1 Configuration Guide ...
Страница 83: ...2 N1259A Connection Guide ...
Страница 107: ...3 N1265A Connection Guide ...
Страница 132: ...3 26 Keysight B1505A Configuration and Connection Guide N1265A Connection Guide Output Connection ...
Страница 133: ...4 N1272A and N1273A Connection Guide ...
Страница 148: ...4 16 Keysight B1505A Configuration and Connection Guide N1272A and N1273A Connection Guide Output Connection ...
Страница 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Страница 203: ...6 Accessory Dimensions ...
Страница 219: ...7 Connection and Ordering Examples ...
Страница 296: ...7 78 Keysight B1505A Configuration and Connection Guide Connection and Ordering Examples Upgrading from existing B1505A ...
Страница 297: ......