7-18
Keysight B1505A Configuration and Connection Guide
Connection and Ordering Examples
Package Device Measurement Configuration Examples
3 kV, 40 A Configuration
•
Device Type: Discrete MOSFET (3 terminals)
•
Key Specification
•
Max voltage/current: 3 kV/40 A without N1259A-300 Module Selector, 3 kV/30 A
with N1259A-300 Module Selector
•
Module Selector for multi-parameter automated measurement function
Figure 7-9
Connection example for 3 kV, 30A automated measurement with Module Selector
HCSMU
Dual HCSMU
Combina
Ɵ
on
adapter
16493S-021
HVSMU
HCSMU
GNDU
MCSMU
B1505A
16493G-001 Digital I/O cable (Included in N1259A-300)
HVSMU2
HCSMU2 F
HCSMU2 S
N1254A-104 TRX(J) to BNC(P) adapter
16493J Interlock cable (included in B1505A)
GNDU2
F
S
F
S
N1259A
F
S
For Gate
For Drain
For Drain
For Drain
HCSMU3 F
HCSMU3 S
16493T (Included in the HVSMU)
16493S (Included in the HCSMU)
16494A (Included in the MCSMU)
16493L (Included in B1505A)
30cm HCSMU cable
(Included in 16493S-021)
16493S (Included in the HCSMU)
Module
Selector
Содержание B1505A
Страница 1: ...Keysight Technologies B1505A Power Device Analyzer Curve Tracer Configuration and Connection Guide ...
Страница 3: ......
Страница 9: ...1 Configuration Guide ...
Страница 83: ...2 N1259A Connection Guide ...
Страница 107: ...3 N1265A Connection Guide ...
Страница 132: ...3 26 Keysight B1505A Configuration and Connection Guide N1265A Connection Guide Output Connection ...
Страница 133: ...4 N1272A and N1273A Connection Guide ...
Страница 148: ...4 16 Keysight B1505A Configuration and Connection Guide N1272A and N1273A Connection Guide Output Connection ...
Страница 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Страница 203: ...6 Accessory Dimensions ...
Страница 219: ...7 Connection and Ordering Examples ...
Страница 296: ...7 78 Keysight B1505A Configuration and Connection Guide Connection and Ordering Examples Upgrading from existing B1505A ...
Страница 297: ......