![Keysight Technologies B1505A Скачать руководство пользователя страница 187](http://html1.mh-extra.com/html/keysight-technologies/b1505a/b1505a_configuration-manual_1961941187.webp)
Keysight B1505A Configuration and Connection Guide
5-39
Connection Guide for Wafer Prober and Your Own Test Fixture
To Connect Ultra High Voltage Expander
Table 5-31
To Connect N1268A/N1269A Output
CAUTION
Never connect the High terminal to any output, including circuit common, chassis ground,
or any other measurement resource such as SMU. Connecting other measurement resource
may damage the connected one.
CAUTION
Never connect the UHVU Low terminal to any other measurement resource except for the
N1269A or N1265A-040 adapter. Connecting other measurement resource may damage
the connected one.
B1505A/N1268A
terminals
Required cables/Description
N1269A input
MCSMU
Force
16494A Triaxial cable, 1.5 m or 3 m
Gate
Force
Sense
16494A Triaxial cable, 1.5 m or 3 m
Sense
MCSMU
Force
16494A Triaxial cable, 1.5 m or 3 m
Chuck
Force
Sense
16494A Triaxial cable, 1.5 m or 3 m
Sense
N1268A Output
Low
SHV(jack)-SHV(jack) cable, 1.5 m or 3 m, of 16493V
Low
N1268A output
Required cables/Description
Connect to
Interlock Output
16493J Interlock cable, 1.5 m or 3 m
Interlock circuit
High
Ultra high voltage coaxial cable with UHV(plug) connector.
UHV cable of N1254A-520 can be used.
DUT high terminal
(ex: drain)
N1269A output
Required cables/Description
Connect to
Gate
Coaxial cable with SHV(plug) connector. SHV cable of
N1254A-520 can be used.
DUT drive terminal
(ex: gate)
Source
Coaxial cable with SHV(plug) connector. SHV cable of
N1254A-520 can be used.
DUT low terminal
(ex: source)
Chuck
Coaxial cable with SHV(plug) connector. SHV cable of
N1254A-520 can be used.
DUT chuck terminal
Содержание B1505A
Страница 1: ...Keysight Technologies B1505A Power Device Analyzer Curve Tracer Configuration and Connection Guide ...
Страница 3: ......
Страница 9: ...1 Configuration Guide ...
Страница 83: ...2 N1259A Connection Guide ...
Страница 107: ...3 N1265A Connection Guide ...
Страница 132: ...3 26 Keysight B1505A Configuration and Connection Guide N1265A Connection Guide Output Connection ...
Страница 133: ...4 N1272A and N1273A Connection Guide ...
Страница 148: ...4 16 Keysight B1505A Configuration and Connection Guide N1272A and N1273A Connection Guide Output Connection ...
Страница 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Страница 203: ...6 Accessory Dimensions ...
Страница 219: ...7 Connection and Ordering Examples ...
Страница 296: ...7 78 Keysight B1505A Configuration and Connection Guide Connection and Ordering Examples Upgrading from existing B1505A ...
Страница 297: ......