30
5) Specifying settings for STEP (Stepped drying) mode
In stepped drying mode it is possible to specify separate measuring conditions for up to a maximum of 5
steps. Specifying a separate drying temperature and separate measuring time for each step makes it
possible to specify finer-grained sets of measuring conditions.
In Step 1, timed halting mode must be specified as the mode used to define halting conditions. In Step 2 and
subsequent steps, it is possible to select between either automatic halting mode or timed halting mode.
*
If automatic halting mode is selected as the mode to be used in Step 2 or later, that step becomes the final step.
Similarly, when timed halting mode is selected and a value of ‘0’ is specified as the measuring time, the previous
step becomes the final step.
Key(s) used
Operating instructions
With the display displaying the weight in grams, press the
key.
If a password has been specified, the message ‘PASS’ will be
displayed to indicate that a password should be entered. Follow
the instructions in “10-2-7 Specifying a password” on p. 46 to
enter the password.
The ‘CONDITIONS’ display area will begin to flash. Press the
key until ‘MODE’ begins to flash.
With the ‘MODE’ display area flashing, press the
key. This
will cause the highlight to shift to measuring mode selection and
for the currently specified measuring mode to begin flashing.
Press the
key until ‘STP’ begins to flash.
*
If ‘STP’ is already flashing, then proceed directly on to the next
step.
When ‘STP’ begins to flash, press the
key. This will cause
‘STP’ to change from a flashing to a constantly lit display, for the
highlight to shift to the specification of the measuring time for Step
1, and for the measuring time to begin to flash.
Press the
key or
key to select the desired measuring
time. The time may be set to a length of anywhere from 1 to 240
minutes in 1 minute increments. Note that here a time of 20
minutes has been specified.
When the desired measuring time flashes, press the
key.
The measuring time will change from a flashing to a constantly lit
display the highlight will shift to the specification of the drying
temperature, and the drying temperature will begin to flash.
Press the
key or
key to select the desired drying
temperature. The temperature may be set to a temperature of
anywhere from 30 to 180
°
C in 1 degree increments. Note that
here a temperature of 105
°
C has been specified.
When the desired drying temperature flashes, press the
key. The step displayed will change from ‘STP 1’ to ‘STP 2’ and
the units area for timed halting mode (‘min’) or automatic halting
mode (‘%’) will begin to flash.
Step
1
2
3
4
5
6
7
8
9
0
Display
Содержание FD-720
Страница 1: ...Infrared Moisture Determination Balance FD 720 Operating Manual ...
Страница 2: ......
Страница 66: ......
Страница 67: ......
Страница 68: ...0706 PA 0201 200 ...