14
6. Assembly and Installation of Main Unit
1
Opening the package
Open the package and check to make sure all listed
items are included.
Scale stopper
Sample dish panel
2
Installing the main unit
• Place the main unit on a flat, stable surface where
it will not be subject to vibrations or exposed to
drafts or breezes.
• Install the main unit so that it is at least 10cm from
the wall both to the front and back and to the left
and right. Make sure there is a space at least 30cm
above the top of the heater cover.
• Do not use it in a closed installation environment.
Depending on the sample, heating may produce
hazardous substances. Install it in a location that
can be adequately ventilated.
4
Checking the power settings
Check the power conversion switch located on the
rear of the main unit to make sure that it is currently
set to the proper voltage setting for your power
source.
* Note that an error will occur if this switch is not set to
a position matching the power source being used.
Power conversion
switch
3
Removing the stopper from the sample dish panel
When using the unit for the first time after purchase,
remove the scale stopper attached below the sample
dish panel as indicated in the accompanying
diagram.
〜
〜
100V
220V
120V
240V
Содержание FD-720
Страница 1: ...Infrared Moisture Determination Balance FD 720 Operating Manual ...
Страница 2: ......
Страница 66: ......
Страница 67: ......
Страница 68: ...0706 PA 0201 200 ...