![KEPCO KLR SERIES Скачать руководство пользователя страница 127](http://html1.mh-extra.com/html/kepco/klr-series/klr-series_developers-manual_1956489127.webp)
KLR-DEV 060713
B-25
KL0
Enables front panel controls.
KL1
Disables front panel controls.
L250
Number (250) of locations programmable using LIST commands.
M40
Number (40) of locations programmable using *SAV and *RCL commands.
SCP
SCPI Programming Language.
B.105 SYSTem:SET? QUERY
SYST:SET?
Syntax
:
Short Form: SYST:SET?
Long Form: SYSTem:SET?
Return Value: <comma separated string>
Description: Returns comma-separated string indicating command functions in effect. See PAR. B.104 for com-
mand function definitions. (See Figure B-7 for example.)
B.106 SYSTem:VERSion? QUERY
SYST:VERS?
Syntax
:
Short Form: SYST:VERS?
Long Form: SYSTem:VERSion?
Return Value: <int_value>.<int_value> (YYYY.V)
Description:
Identifies SCPI Version implemented.
Returns SCPI Version number:
YYYY = year, V = Revision number for specified year. (See example, Figure B-7.)
B.107 TRIGger[:SEQuence]:SOURce COMMAND
TRIG:SOUR
Syntax
:
Short Form: TRIG[:SEQ]:SOUR {EXT | BUS | IMM}
Long Form: TRIGger[:SEQuence]:SOURce {EXT | BUS | IMM}
Description:
Selects the active trigger source.
When TRIG:SOUR BUS is set, an event trigger is created by
sending either *TRG or GPIB <GET> commands. When TRIG:SOUR EXT is set, the trigger is created
by connecting the external trigger line (J2, pin 4 or pin 3 of the mating connector for J2) to analog sig-
nal ground (J2, pin 1 or J2 mating connector, pin 1). The trigger must be armed by sending either
INIT:IMM or INIT:CONT enable (ON) commands prior to sending the selected trigger command. (see
PAR. A.16). Receipt of the active trigger causes the power supply to program to the previously stored
values of VOLT:TRIG and CURR:TRIG (actual output voltage/current is dependent upon the load).
Sending TRIG:SOUR IMM causes the next VOLT:TRIG or CURR:TRIG to immediately program the
output, regardless of trigger arming. *RST sets TRIG:SOUR to IMMediate. Related Commands:
<GET>, *TRG, INIT:IMM, INIT:CONT.
B.108 TRIGger[:SEQuence]:SOURce? QUERY
TRIG:SOUR?
Syntax
:
Short Form: TRIG[:SEQ]:SOUR?
Long Form: TRIGger[:SEQuence]:SOURce?
Return Value: Return Value: <string> EXT or BUS or IMM
Description:
Returns string indicating trigger source.
EXT indicates trigger source is J2 pin 14. BUS indicates
source is either *TRG or GPIB <GET> command IMM indicates source is VOLT:TRIG or CURR:TRIG
command. See PAR. B.107 for details.
TABLE B-5. ERROR MESSAGES
ERROR MESSAGE
ESR ERROR BIT SET
(SEE PAR. A.5)
EXPLANATION
0,“No error”
None
No error
-100,“Command error”
Command Error bit 5
Command and data understood, but more information included which
is not recognized.
-102,”Syntax error”
Command Error bit 5 First 4 characters recognized, subsequent characters not recognized.
-103,”Invalid separator”
Command Error bit 5 For example, VOLT.PROT received instead of VOLT:PROT.
-108,”Parameter Not Allowed Error”
Command Error bit 5 Volt12 sequence, channel number is invalid.
-109,”Missing parameter”
Command Error bit 5 For example, VOLT instead of VOLT 21.
-111,”Header separator error”
Command Error bit 5 Missing space between volt and value or ; missing.
-113,”Undefined header”
Command Error bit 5
First 4 characters could not be identified as legal command.For
example, command VLT instead of VOLT.
-120,”Numeric data error”
Command Error bit 5 Expected number but other characters were detected.
-121,”Invalid character in number”
Command Error bit 5 Volt 1,500 (comma not allowed).
Содержание KLR SERIES
Страница 2: ......
Страница 10: ......
Страница 42: ......
Страница 58: ...4 16 KLR DEV 060713 FIGURE 4 10 FUNCTION GENERATOR PANEL FIGURE 4 11 CONFIGURE User Sequence vi BLOCK DIAGRAM ...
Страница 59: ......
Страница 60: ......
Страница 61: ...FIGURE 4 13 FUNCTION GE ...
Страница 62: ......
Страница 86: ......
Страница 90: ......