![KEPCO KLR SERIES Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/kepco/klr-series/klr-series_developers-manual_1956489057.webp)
KLR-DEV 060713
4-15
4.6.4
FUNCTION GENERATOR (Supported in Standalone Configurations Only)
NOTE: Complex, dynamic waveshapes typically require a power supply output that can both
source and sink current, however KLR models without R Option can only source cur-
rent, and depend on the load to sink current. KLR Models with R Option can sink cur-
rent for brief periods.
When the Function Generator button on the Front Panel is clicked, the function generator panel
(Figure 4-10) is displayed. The frequency slider determines the frequency of the generated
waveform. The number of points determines the limits for frequency adjustment. For example, if
10 points are selected, the maximum frequency is 100 Hz, the lowest frequency is 0.0015 Hz.
The number of points determines the smoothness of the waveform because it is constructed
incrementally, but it also limits the maximum frequency. For example, if the 250 (maximum)
points are selected, the frequency is 0.00001 Hz.
The Amplitude and Offset sliders are to the right of the Frequency slider. When the requested
output exceeds the capability of the KLR, the output is clipped by the function generator. Below
these sliders is the Waveform Type dropdown selector, either Sine, Sawtooth, Triangle or
Square. Below the Waveform Type dropdown is a visual display of the points that have been
specified. To the left of the displayed waveform is the Duty Cycle slider which affects square
wave-shapes only. The Phase slider shifts the start point of the waveform. As these sliders are
changed, the digital display beneath the slider shows the setting. Changes to the slider settings
immediately change the Commanded Waveform display to reflect the new parameters.
Above the Start button, two additional parameters required for the execution of the waveform
are displayed. The value of the non-changing parameter, current (in Amperes) for a voltage
waveform (shown in Figure 4-13) or voltage (in Volts) for a current waveform must be specified.
The # of Cycles specifies how many waveforms to produce; if set to 0 the waveform repeats
continuously.
Once the Start button is clicked, the commands are sequentially sent to the KLR to build the
waveform by incrementally changing the output. While the waveform is being generated, peri-
odic measurements of the KLR output are made and shown in the Measured Waveform display.
To stop the waveform click on the Stop button, which is red while the waveform is running.
Figure 4-13 is a block diagram that shows how the function generator was implemented using
list commands in this example. The
KepcoDCPwr Clear All Lists.vi
was executed first. This VI
clears the previous list, allowing the new list to be sent to the KLR. Lists are not deleted from the
KLR except when a limit model setting is changed or when the Clear All function is executed.
The
KepcoDCPwr Configure User Sequence.vi
does the actual downloading of the points to the
KLR unit. The
KepcoDCPwr Configure Voltage Mode.vi
causes the list to be executed.
The
KepcoDCPwr Configure User Sequence.vi
is shown in Figure 4-11. This logic is specifically
configured to generate a voltage list where two arrays of voltage and dwell times are supplied,
along with the operating current and number of cycles for the list to execute. In the left-most
case statement there is a command to send one entry for the control and current lists of the
internal arbitrary waveform generator. The last two functions in the diagram send the dwell and
then the voltage list to the KLR. These functions, along with the individual write of the
LIST:COUN command, set up the arbitrary waveform generator for the selected number of rep-
etitions.
Содержание KLR SERIES
Страница 2: ......
Страница 10: ......
Страница 42: ......
Страница 58: ...4 16 KLR DEV 060713 FIGURE 4 10 FUNCTION GENERATOR PANEL FIGURE 4 11 CONFIGURE User Sequence vi BLOCK DIAGRAM ...
Страница 59: ......
Страница 60: ......
Страница 61: ...FIGURE 4 13 FUNCTION GE ...
Страница 62: ......
Страница 86: ......
Страница 90: ......