KLR-DEV 060713
B-3
FIGURE B-1. PROGRAMMING THE OUTPUT
B.7
INITiate:CONTinuous COMMAND
INIT:CONT
Syntax:
Short Form: INIT:CONT {ON | OFF} or {1 | 0} (1 = on, 0 = off)
Long Form: INITiate:CONTinuous {ON | OFF} or {1 | 0} (1 = on, 0 = off)
Default setting is OFF.
Description:
INIT:CONT ON enables continuous triggers.; INIT:CONT OFF disables continuous triggers.
If
INIT:CONT is OFF, then INIT[:IMM] arms the trigger system for a single trigger. If INIT:CONT is ON,
then the trigger system is continuously armed. INIT:CONT ON sets the WTG bit in the Questionable
Status register to 1 after each trigger event. Executing *RST command sets INIT:CONT to OFF. (See
example, Figure B-1.)
NOTE: The following example assumes KLR 75-32 with limit model set for 75V, 16A and output operating in volt-
age stabilization mode. See Figure B-3 for directions on programming limit model settings.
OUTP ON
Enables output.
OUTP?
Returns 1 (output enabled).
STAT:PRES
Disables all status event reporting.
STAT:OPER:ENAB 32
Enables only bit 5 (WTG, Waiting for Trigger)
STAT:OPER:ENAB?
Returns 32, WTG bit enabled.
VOLT 75;CURR 5E-1
Programs output voltage to 75V, current limit to 0.5A.
INIT:CONT ON
Enables continuous triggering.
INIT:CONT?
Returns 1 (continuous triggering enabled).
STAT:QUES?
Returns 32, indicating WTG bit is set (trigger armed).
VOLT:TRIG 31.5;CURR:TRIG8E-1 Programs output voltage for 31.5V, current limit for 0.8A upon receipt of trigger
stimulus.
*TRG
Output voltage is set to 31.5V with 0.8A current limit.
STAT:QUES?
Returns 32 again, indicating WTG bit was set again (continuous triggering).
VOLT 32.1;CURR 5
Programs output voltage to 32.1V, current limit to 5A.
MEAS:VOLT?
Returns actual output voltage (if output voltage is 32.15V, power supply
returns 3.215E1).
MEAS:CURR?
Returns actual output current.
FUNC:MODE?
Returns operating mode of power supply followed by programmed operating
mode, e.g., VOLT,VOLT.
CURR:TRIG?
Returns 5E0 (current value established by CURR:TRIG).
VOLT:TRIG?
Returns 3.15E1 (voltage value established by VOLT:TRIG).
ABOR
Changes pending trigger levels to immediate values (32.1V, 5A).
VOLT 37.7;CURR 2.5
Programs output voltage to 37.7V, current limit to 2.5A.
*TRG
Output returns to trigger values established by ABOR (32.1V, 5A).
INIT:CONT 0
Disables continuous triggering.
INIT:CONT?
Returns 0 (continuous triggering disabled).
INIT:IMM
Arms single trigger event.
STAT:QUES?
Returns 32 indicating WTG bit is set (trigger armed).
*TRG
Output returns to trigger values established by ABOR (32.1V, 5A).
STAT:QUES?
Returns 0, indicating trigger is not rearmed (single trigger event).
VOLT 37.7, CURR 2.5
Programs output voltage to 37.7V, current limit to 2.5A.
*TRG
Output remains in current state as init:cont is set to 1.
INIT
Trigger is armed for a single event
*TRG
Output returns to previous trigger values (32.1V, 5A) without rearming trigger.
OUTP OFF
Output disabled.
OUTP?
Returns 0 (output disabled).
MEAS:VOLT?
Returns 0 (measured output voltage).
VOLT?
Returns 3.21E1 (last programmed output voltage).
CURR?
Returns 5E0 (last programmed output current).
CURR? MAX
Returns 1.6E1 (maximum allowable current for present limit model).
CURR? MIN
Returns 4E-1 (minimum allowable programming current for model).
Содержание KLR SERIES
Страница 2: ......
Страница 10: ......
Страница 42: ......
Страница 58: ...4 16 KLR DEV 060713 FIGURE 4 10 FUNCTION GENERATOR PANEL FIGURE 4 11 CONFIGURE User Sequence vi BLOCK DIAGRAM ...
Страница 59: ......
Страница 60: ......
Страница 61: ...FIGURE 4 13 FUNCTION GE ...
Страница 62: ......
Страница 86: ......
Страница 90: ......