B-18
KLR-DEV 060713
B.64 STATus:QUEStionable:CONDition? QUERY
STAT:QUES:COND?
Syntax
:
Short Form: STAT:QUES:COND?
Long Form: STATus:QUEStionable:CONDition?
Return Value: <int_value> actual register value
Description:
Returns the value of the Questionable Condition Register (see Table B-3).
The Questionable
Condition Register contains unlatched real-time information about questionable conditions of the
power supply. Bit set to 1 = condition (active, true); bit reset to 0 = condition (inactive, false). (See
example, Figure B-6.)
B.65 STATus:QUEStionable:ENABle COMMAND
STAT:QUES:ENAB
Syntax
:
Short Form: STAT:QUES:ENAB <int_value>
Long Form: STATus:QUESionable:ENABle <int_value>
Function:
Programs Questionable Condition Enable Register.
Description:
Programs Questionable Condition Enable Register (see Table B-3).
The Questionable Condition
Enable Register determines which conditions are allowed to set the Questionable Condition Register;
it is a mask for enabling
specific bits in the Questionable Event register that can cause the question-
able summary bit (bit 3) of the Status Byte register to be set. The questionable summary bit is the log-
ical OR of all the enabled bits in the Questionable Event register
.
Bit set to 1 = function enabled
(active, true); bit reset to 0 = function disabled (inactive, false)
.
(See example, Figure B-6.)
B.66 STATus:QUEStionable:ENABle? QUERY
STAT:QUES:ENAB?
Syntax
:
Short Form: STAT:QUES:ENAB?
Long Form: STATus:QUEStionable:ENABle?
Return Value: <int_value> actual register value
Description:
Reads Questionable Condition Enable Register (see Table B-3).
Power supply returns value of
Questionable Condition Enable Register, indicating which conditions are being monitored. Bit set to 1
= function enabled (active, true); bit reset to 0 = function disabled (inactive, false)
.
Related Com-
mands: STAT:QUES?. (See example, Figure B-6.)
B.67 SYSTem:COMMunication:GPIB:ADDRess COMMAND
SYST:COMM:GPIB:ADDR
Syntax
:
Short Form: SYST:COMM:GPIB:ADDR<int_value> 1 to 30
Long Form: SYSTem:COMMunication:GPIB:ADDRess<int_value> 01 to 30
Description: Sets selected power supply GPIB address.
B.68 SYSTem:COMMunication:GPIB:ADDRess? QUERY
SYST:COMM:GPIB:ADDR?
Syntax
:
Short Form: SYST:COMM:GPIB:ADDR?
Long Form: SYSTem:COMMunication:GPIB:ADDRess?
Return Value: <int_value> (GPIB address)
Description:
Returns GPIB address.
B.69 SYSTem:COMMunication:LAN:AUTO COMMAND
SYST:COMM:LAN:AUTO
[E-SERIES MODELS ONLY]
Syntax
:
Short Form: SYST:COMM:LAN:AUTO {ON | OFF}
Long Form: SYSTem:COMMunication:LAN:AUTO {ON | OFF}
Description:
Specifies the AUTO MODE.
When set to OFF, the static addresses set by the user from the front
panel or using SYST:COMM:LAN commands are used for LAN communication. When set to ON, the
unit obtains a valid IP address automatically.
TABLE B-3. QUESTIONABLE EVENT REGISTER, QUESTIONABLE CONDITION REGISTER
AND QUESTIONABLE CONDITION ENABLE REGISTER BITS
CONDITION
NU
M/S
FAN
PWR
OTP
OLF
OCP
OVP
BIT
15 - 7
6
5
4
3
2
1
0
VALUE
32,768 - 128
64
32
16
8
4
2
1
M/S - MASTER/SLAVE FAILURE
FAN - INTERNAL FAN FAILURE
PWR - LOSS OF SOURCE POWER
OTP - OVERTEMPERATURE
OLF - OUTPUT LEAD FAULT
OCP - OVERCURRENT
OVP - OVERVOLTAGE
NU - NOT USED
Содержание KLR SERIES
Страница 2: ......
Страница 10: ......
Страница 42: ......
Страница 58: ...4 16 KLR DEV 060713 FIGURE 4 10 FUNCTION GENERATOR PANEL FIGURE 4 11 CONFIGURE User Sequence vi BLOCK DIAGRAM ...
Страница 59: ......
Страница 60: ......
Страница 61: ...FIGURE 4 13 FUNCTION GE ...
Страница 62: ......
Страница 86: ......
Страница 90: ......