Performance Verification
1-3
Environmental conditions
Conduct your performance verification procedures in a test environment with:
• An ambient temperature of 18 to 28°C (65 to 82°F).
• A relative humidity of less than 70% unless otherwise noted.
Warm-up period
Allow the Model 2425 to warm up for at least one hour before conducting the verification
procedures.
If the instrument has been subjected to temperature extremes (those outside the ranges
stated above), allow additional time for the instrument’s internal temperature to stabilize.
Typically, allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified
temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the manu-
facturer.
Line power
The Model 2425 requires a line voltage of 100 to 240V and a line frequency of 50 or
60Hz. Verification tests must be performed within this range.
Содержание SourceMeter 2425
Страница 1: ...Model 2425100W SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031...
Страница 15: ......
Страница 16: ...1 PerformanceVerification...
Страница 35: ...1 20 Performance Verification Model 2425 Service Manual...
Страница 36: ...2 Calibration...
Страница 62: ...3 RoutineMaintenance...
Страница 65: ...3 4 Routine Maintenance Model 2425 Service Manual...
Страница 66: ...4 Troubleshooting...
Страница 84: ...5 Disassembly...
Страница 92: ...6 ReplacementParts...
Страница 113: ...6 22 Replacement Parts Model 2425 Service Manual...
Страница 114: ...A Specifications...
Страница 116: ...B CommandReference...
Страница 128: ...C CalibrationPrograms...
Страница 139: ......
Страница 141: ......